ISO 5347-7:1993Withdrawn

Methods for the calibration of vibration and shock pick-ups — Part 7: Primary calibration by centrifuge

Méthodes pour l'étalonnage de capteurs de vibrations et de chocs — Partie 7: Étalonnage primaire par centrifugeur

This edition has been replaced by ISO 16063-17:2016.

Abstract / scope

Lays down detailed specifications for the instrumentation and procedure to be used for primary calibration of accelerometers using centrifuge calibration. Applies to rectilinear accelerometers with zero-frequency response, mainly of the strain gauge or piezoresistive type, and to primary standard and working pick-ups. Is applicable for a calibration range from 10 m/s^2 to 1000 m/s^2 (higher accelerations possible) at 0 Hz. The limits of uncertainty applicable are ± 1 % of reading.

Life cycle

Replaced by

ISO 16063-17:2016PublishedMethods for the calibration of vibration and shock transducers — Part 17: Primary calibration by centrifugeInternational Standard · Published 30 May 2016 · ISO/TC 108