ISO 11438-4:1993Withdrawn

Ferronickel — Determination of trace-element content by electrothermal atomic absorption spectrometric method — Part 4: Determination of tin content

Ferro-nickel — Dosage des éléments-traces — Méthode par spectrométrie d'absorption atomique à excitation électrothermique — Partie 4: Dosage de l'étain

Abstract / scope

The principle of the method specified is dissolution of a test portion in nitric acid, measurement of the absorption of the resonance line energy from the spectrum of tin in the test solution at a wavelength of 286,3 nm by an atomic absorption spectrometer fitted with a graphite furnace electrothermal atomizer, calibration by the standard additions method described in ISO 11438-1. Applies to tin contents in the range of 1,0 g/t to 10,0 g/t in ferronickel.