ISO/ASTM 51631:2002Withdrawn

Practice for use of calorimetric dosimetry systems for electron beam dose measurements and dosimeter calibrations

Pratique de l'utilisation des systèmes dosimétriques calorimétriques pour des mesures de dose délivrée par un faisceau d'électrons et pour l'étalonnage de dosimètres

Abstract / scope

ISO/ASTM 51631 covers the preparation and use of semi-adiabatic calorimeters for measurement of absorbed doses in graphite, water or polystyrene when irradiated with electrons. The calorimeters are either transported by a conveyor past a scanned electron beam or are stationary in a broadened beam. It also covers the use of these calorimeters to calibrate dosimeter systems in electron beams intended for radiation processing applications.

This International Standard applies to electron beams in the energy range from 4 MeV to 12 MeV.

The absorbed dose range depends on the absorbing material and the irradiation and measurement conditions. The minimum dose is approximately 100 Gy and maximum dose is approximately 50 kGy.

The averaged absorbed dose rate range shall generally be greater than 10 Gy·s-1 , but depends on the same conditions as above.

The temperature range for use of these calorimeters depends on the thermal resistance of the materials and on the calibration range of the temperature sensor.

Life cycle

Replaces

ISO 15568:1998WithdrawnPractice for use of calorimetric dosimetry systems for electron beam dose measurements and dosimeter calibrationsInternational Standard · Published 20 Dec 1998 · ISO/TC 85

Replaced by

ISO/ASTM 51631:2003WithdrawnPractice for use of calorimetric dosimetry systems for electron beam dose measurements and dosimeter calibrationsInternational Standard · Published 18 Aug 2003 · ISO/TC 85