ISO/ASTM 51540:2002Withdrawn

Practice for use of a radiochromic liquid dosimetry system

Pratique de l'utilisation d'un système dosimétrique radiochromique liquide

Abstract / scope

ISO/ASTM 51540 covers the preparation, handling, testing and procedure for using radiochromic liquid dosimetry systems of radiochromic dye solutions held in sealed or capped containers (e.g., ampoules, vials). It also covers the use of spectrophotometric or photometric read-out equipment for measuring absorbed doses in materials irradiated by photons and electrons.

This practice applies to radiochromic liquid dosimeter solutions that can be used within part or all of the specified ranges as follows:

the absorbed dose range is from 0,5 Gy to 40 000 Gy for photons and electrons;

the absorbed dose rate is from 10-3 Gy s -1to 1011 Gy s-1;

the radiation energy range for photons is from 0,01 MeV to 20 MeV;

the radiation energy range for electrons is from 0,01 MeV to 20 MeV;

the irradiation temperature range is from - 40 °C to + 60 °C.

Life cycle

Replaces

ISO 15565:1998WithdrawnPractice for use of a radiochromic liquid dosimetry systemInternational Standard · Published 20 Dec 1998 · ISO/TC 85

Replaced by

ISO/ASTM 51540:2004WithdrawnPractice for use of a radiochromic liquid dosimetry systemInternational Standard · Published 8 Jul 2004 · ISO/TC 85