ISO/ASTM 51607:2002Withdrawn

Practice for use of the alanine-EPR dosimetriy system

Pratique de l'utilisation d'un système dosimétrique à l'alanine utilisant la résonance paramagnétique électronique

Abstract / scope

ISO/ASTM 51607 covers materials description, dosimeter preparation, instrumentation and procedures for using the alanine-EPR dosimetry system for measuring the absorbed dose in materials irradiated with photons and electrons. The system is based on electron paramagnetic resonance (EPR) spectroscopy of free radicals derived from the amino acid alanine. It is classified as a reference standard dosimetry system.

This International Standard covers alanine-EPR dosimetry systems for dose measurements under the following conditions:

the absorbed dose range is between 1 Gy and 105 Gy;

the absorbed dose rate is up to 102 Gy s -1 for continuous radiation fields and up to 5 times 107 Gy s-1 for pulsed radiation fields;

the radiation energy for photons and electrons is between 0,1 MeV and 28 MeV;

the irradiation temperature is between - 60 °C and + 90 °C.

Life cycle

Replaces

ISO 15566:1998WithdrawnPractice for use of the alanine-EPR dosimetriy systemInternational Standard · Published 20 Dec 1998 · ISO/TC 85

Replaced by

ISO/ASTM 51607:2004WithdrawnPractice for use of the alanine-EPR dosimetry systemInternational Standard · Published 25 Oct 2004 · ISO/TC 85