ISO 27186:2010Withdrawn

Active implantable medical devices — Four-pole connector system for implantable cardiac rhythm management devices - Dimensional and test requirements

Dispositifs médicaux actifs implantables — Systèmes de branchement à quatre pôles pour gérer le rhythme cardiaque — Dimensions et exigences d'essai

Abstract / scope

ISO 27186:2010 specifies a four-pole connector system for implantable cardiac rhythm management devices which have pacing, electrogram sensing and/or defibrillation functions. This International Standard includes requirements for the connector portion of an implantable lead as well as for the mating connector cavity attached to an implantable pulse generator. Essential dimensions and performance requirements are specified together with appropriate test methods.

Life cycle

Replaced by

ISO 27186:2020WithdrawnActive implantable medical devices — Four-pole connector system for implantable cardiac rhythm management devices — Dimensional and test requirementsInternational Standard · Published 30 Nov 2020 · ISO/TC 150/SC 6