ISO 29301:2010Withdrawn

Microbeam analysis — Analytical transmission electron microscopy — Methods for calibrating image magnification by using reference materials having periodic structures

Analyse par microfaisceaux — Microscopie électronique en transmission analytique — Méthodes d'étalonnage du grandissement d'image au moyen de matériaux de référence de structures périodiques

Abstract / scope

ISO 29301:2010 specifies a calibration procedure applicable to images recorded over a wide magnification range in a transmission electron microscope (TEM). The reference materials used for calibration possess a periodic structure, such as a diffraction grating replica, a super-lattice structure of semiconductor or an analysing crystal for X-ray analysis, and a crystal lattice image of carbon, gold or silicon. ISO 29301:2010 is applicable to the magnification of the TEM image recorded on a photographic film, or an imaging plate, or detected by an image sensor built into a digital camera. ISO 29301:2010 also refers to the calibration of a scale bar. ISO 29301:2010 does not apply to the dedicated critical dimension measurement TEM (CD-TEM) and the scanning transmission electron microscope (STEM)

Life cycle

Replaced by

ISO 29301:2017WithdrawnMicrobeam analysis — Analytical electron microscopy — Methods for calibrating image magnification by using reference materials with periodic structuresInternational Standard · Published 6 Dec 2017 · ISO/TC 202/SC 3