ISO 15883-6:2011Withdrawn

Washer-disinfectors — Part 6: Requirements and tests for washer-disinfectors employing thermal disinfection for non-invasive, non-critical medical devices and healthcare equipment

Laveurs désinfecteurs — Partie 6: Exigences et essais pour les laveurs désinfecteurs utilisant une désinfection thermique pour les dispositifs médicaux non invasifs, non critiques et pour l'équipement de soins de santé

This edition has been replaced by ISO 15883-6:2026.

Abstract / scope

ISO 15883-6:2011 specifies particular requirements for washer-disinfectors (WDs) intended for use when the level of assurance of disinfection that is necessary can be achieved by cleaning and thermal disinfection (A0 not less than 60) and does not require an independent automated record of critical processes to be kept. It is intended to be used in conjunction with ISO 15883-1, which gives general requirements for WDs.

The range of products on which WDs of this particular type can be used is restricted to devices and equipment which are non-invasive and non-critical (i.e. not penetrating skin or contacting mucosal surfaces).

Life cycle

Replaced by

ISO 15883-6:2026PublishedWasher-disinfectors — Part 6: Requirements and tests for washer-disinfectors employing thermal disinfection for non-critical medical devices and health care equipmentInternational Standard · Published 26 Jan 2026 · ISO/TC 198