ISO 17378-2:2014Published

Water quality — Determination of arsenic and antimony — Part 2: Method using hydride generation atomic absorption spectrometry (HG-AAS)

Qualité de l'eau — Dosage de l'arsenic et de l'antimoine — Partie 2: Méthode par spectrométrie d'absorption atomique à génération d'hydrures (HG-AAS)

Abstract / scope

ISO 17378-2:2014 specifies a method for the determination of arsenic and antimony. The method is applicable to drinking water, surface water, ground water, and rain water. The approximate linear application range of ISO 17378-2:2014 for both elements is from 0,5 µg/l to 20 µg/l. Samples containing higher concentrations than the application range can be analysed following appropriate dilution.

Generally sea water is outside the scope of ISO 17378-2:2014. Sea water samples can be analysed using a standard additions approach providing that this is validated for the samples under test. The method is unlikely to detect organo-arsenic and organo-antimony compounds.

The sensitivity of this method is dependent on the selected operating conditions.

Life cycle

Replaces

ISO 11969:1996WithdrawnWater quality — Determination of arsenic — Atomic absorption spectrometric method (hydride technique)International Standard · Published 11 Jul 1996 · ISO/TC 147/SC 2