Inhalational anaesthesia systems — Draw-over anaesthetic systems

Systèmes d'anesthésie par inhalation — Alimentation en vapeur et équipements annexes

Abstract / scope

ISO 18835:2015 specifies basic safety and essential performance requirements for anaesthetic systems utilizing the draw-over method to provide inhalational anaesthesia.

Requirements are included to allow the use of these systems with both non-flammable and flammable anaesthetic agents.

ISO 18835:2015 also includes requirements for a bellows-type manual ventilator.

NOTE 1 Requirements for automatic anaesthetic ventilators are covered by ISO 80601‑2-13.

NOTE 2 Requirements for operator-powered self-inflating bags are covered by ISO 10651‑4.

ISO 18835:2015 does not specify requirements for monitoring of the equipment or the patient.

Life cycle

Replaces

ISO/TS 18835:2004WithdrawnInhalational anaesthesia systems — Draw-over vaporizers and associated equipmentTechnical Specification · Published 2 Jul 2004 · ISO/TC 121/SC 1