ISO/ASTM 51205:2017Withdrawn

Practice for use of a ceric-cerous sulfate dosimetry system

Pratique de l'utilisation d'un système dosimétrique au sulfate cérique-céreux

Abstract / scope

1.1 This practice covers the preparation, testing, and procedure for using the ceric-cerous sulfate dosimetry system to measure absorbed dose to water when exposed to ionizing radiation. The system consists of a dosimeter and appropriate analytical instrumentation. For simplicity, the system will be referred to as the ceric-cerous system. The ceric-cerous dosimeter is classified as a type 1 dosimeter on the basis of the effect of influence quantities. The ceric-cerous system may be used as a reference standard dosimetry system or as a routine dosimetry system.

1.2 ISO/ASTM 51205:2017 is one of a set of standards that provides recommendations for properly implementing dosimetry in radiation processing, and describes a means of achieving compliance with the requirements of ISO/ASTM Practice 52628 for the ceric-cerous system. It is intended to be read in conjunction with ISO/ASTM Practice 52628.

1.3 This practice describes both the spectrophotometric and the potentiometric readout procedures for the ceric-cerous system.

1.4 This practice applies only to gamma radiation, X-radiation/bremsstrahlung, and high energy electrons.

Life cycle

Replaces

ISO/ASTM 51205:2009WithdrawnPractice for use of a ceric-cerous sulfate dosimetry systemInternational Standard · Published 7 Jul 2009 · ISO/TC 85