ISO 13084:2018Withdrawn

Surface chemical analysis — Secondary ion mass spectrometry — Calibration of the mass scale for a time-of-flight secondary ion mass spectrometer

Analyse chimique des surfaces — Spectrométrie de masse des ions secondaires — Étalonnage de l'échelle de masse pour un spectromètre de masse des ions secondaires à temps de vol

This edition has been replaced by ISO 13084:2025.

Abstract / scope

This document specifies a method to optimize the mass calibration accuracy in time-of-flight secondary ion mass spectrometry (SIMS) instruments used for general analytical purposes. It is only applicable to time-of-flight instruments but is not restricted to any particular instrument design. Guidance is provided for some of the instrumental parameters that can be optimized using this procedure and the types of generic peaks suitable to calibrate the mass scale for optimum mass accuracy.

Life cycle

Replaces

ISO 13084:2011WithdrawnSurface chemical analysis — Secondary-ion mass spectrometry — Calibration of the mass scale for a time-of-flight secondary-ion mass spectrometerInternational Standard · Published 5 May 2011 · ISO/TC 201/SC 6

Replaced by

ISO 13084:2025PublishedSurface chemical analysis — Mass spectrometries — Calibration of the mass scale for a time-of-flight secondary ion mass spectrometerInternational Standard · Published 23 Oct 2025 · ISO/TC 201/SC 6