ISO 11608-5:2022Published

Needle-based injection systems for medical use — Requirements and test methods — Part 5: Automated functions

Systèmes d'injection à aiguille pour usage médical — Exigences et méthodes d'essai — Partie 5: Fonctions automatisées

Abstract / scope

This document specifies requirements and test methods for automated functions in needle-based injection systems with automated functions (NIS-AUTO).

General requirements are provided for all automated functions. In addition, specific requirements are provided for the following automated functions:

a) medicinal product preparation (e.g. reconstitution);

b) needle preparation;

c) needle hiding;

d) priming;

e) dose setting;

f) needle insertion;

g) injection depth control;

h) injection of the medicinal product;

i) recording of device functions;

NOTE  This document does not cover remote communication from the NIS-AUTO (pertains to wired and wireless communication transfer from the NIS auto).

j) disabling the NIS-AUTO;

k) needle retraction;

l) needle shielding;

m) needle removal.

All references to "function" in this document are by definition construed as automated functions (see 3.2). This document does not apply to functions that are performed manually by the user.

Life cycle

Replaces

ISO 11608-5:2012WithdrawnNeedle-based injection systems for medical use — Requirements and test methods — Part 5: Automated functionsInternational Standard · Published 20 Sep 2012 · ISO/TC 84