ISO 10993-7:2026Published

Biological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals

Évaluation biologique des dispositifs médicaux — Partie 7: Résidus de stérilisation à l'oxyde d'éthylène

Abstract / scope

This document specifies allowable limits (AL) for residual ethylene oxide (EO) and ethylene chlorohydrin (ECH) in EO-sterilized medical devices, procedures for the measurement of EO and ECH, and methods for determining conformity so that devices can be released. Additional background, including guidance and a flowchart showing how this document is applied, are also included in Annexes A, B, C, D, E, F, G, H, I, J and K.

EO-sterilized devices or components that have neither direct nor indirect body or user contact (e.g. in vitro diagnostic devices) are out of scope of this document. This document does not apply to devices that have been demonstrated to not absorb or retain EO or its degradation product ECH, such as medical devices made exclusively of metal alloys and glass, see Clause C.5[228].

NOTE            This document does not specify limits for ethylene glycol (EG). No device limits are specified for EG because the risk assessment in Annex F indicates that calculated allowable levels are higher than those likely to occur in a medical device.

Life cycle

Replaces

ISO 10993-7:2008WithdrawnBiological evaluation of medical devices — Part 7: Ethylene oxide sterilization residualsInternational Standard · Published 13 Oct 2008 · ISO/TC 194ISO 10993-7:2008/Cor 1:2009WithdrawnBiological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals — Technical Corrigendum 1International Standard · Published 12 Nov 2009 · ISO/TC 194ISO 10993-7:2008/Amd 1:2019WithdrawnBiological evaluation of medical devices — Part 7: Ethylene oxide sterilization residuals — Amendment 1: Applicability of allowable limits for neonates and infantsInternational Standard · Published 9 Dec 2019 · ISO/TC 194