ISO/PWI 16063-46.2Under development

Methods for the calibration of vibration and shock transducers — Part 46: Part 46: Low-frequency calibration by machine vision systems

Titre manque — Partie 46: Titre manque

Abstract / scope

This document specifies the instrumentation and procedure used for vibration calibration of rectilinear vibration transducer (with or without amplifier) to obtain the low-frequency complex sensitivity by machine vision systems. It is applicable to a frequency range from 0,01 Hz to 20 Hz and a dynamic range (amplitude) from 0,000 4 m/s2 (200 mm displacement peak-peak at 0,01 Hz) to 20 m/s2.
NOTE: The lower limit of frequency range depends on the noise of the vibration transducer and the stroke of the vibrator. The lower limit of dynamic range depends on the noise of the vibration transducer and environmental noise.